Operation Principles
Cell Illumination
3
Figure 3.2
Laser Beam Shaping
Cell Illumination
As cells in the sample stream go through the sensing area of the flow cell, the elliptical laser beam
illuminates them. The cells scatter the laser light and emit fluorescent light from fluorescent dyes
attached to them.
Forward Scatter
The amount of laser light scattered at narrow angles to the axis of the laser beam is called forward
scatter (FS). The amount of FS is proportional to the size of the cell that scattered the laser light.
Side Scatter and Fluorescent Light
The amount of laser light scattered at about a 90° angle to the axis of the laser beam is called side
scatter (SS). The amount of SS is proportional to the granularity of the cell that scattered the laser
light. For example, SS is used to differentiate between lymphocytes, monocytes, and granulocytes.
In addition to the SS, the cells emit fluorescent light (FL) at all angles to the axis of the laser beam.
The amount of FL enables the instrument to measure characteristics of the cells emitting that light,
depending on the reagents used. For example, FL is used to identify molecules, such as cell surface
antigens.
1.
Violet laser beams
2.
Blue laser beam
3.
Red laser beam
4.
Horizontal beam shaping lens
5.
Vertical beam shaping lens
b
c
d
e
f
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......