Specifications
Performance Characteristics And Specifications
Scatter Resolution
Scatter resolution on the Navios Flow Cytometer was measured using 0.404
μ
m particles from
Thermo Scientific. The noise was set to the bottom of the scale, using FS HV = 700, Gain at 200,
Discriminator at 47 and setting the Forward Scatter detector to the Wide2 setting. The Navios was
able to clearly show baseline resolution between the noise and the 0.404
μ
m diameter particles.
Forward Scatter
The HPCV of the integral signal intensity values using Flow-Check Pro fluorospheres is <2% from the
blue laser.
Fluorescence
The HPCV of the integral signal intensity values using Flow-Check Pro fluorospheres is <2% for FL1-
FL4 and <2.5% for FL5 from the blue laser, <3.0% for FL6-FL8 from the red laser and <4.0% for FL9-
FL10 from the violet laser.
Sensitivity
Fluorescence
Three measurements of multi-level fluorescence sensitivity particles were taken on 3 Navios Flow
Cytometers. Acquisition was conducted on 10,000 bead stop count at the medium flow rate. For the
FITC and PE measurements, 8 peak Spherotech Rainbow Calibration Particles (RCP-30-5A) were
used. For the PeCY5, and APC, Spherotech Ultra Rainbow Calibration Particles (URCP-38-2K) were
used. The following values represent the average of the measurements taken.
•
<112 MESF for FITC
•
<78 MESF for PE
•
<15 MESF for PC5
•
<75 MESF for APC
Laser Characteristics
Day-To-Day
The mean channel value of the integral signal intensity of alignment verification fluorospheres
from the standard PMT’s off the blue, red and violet lasers does not vary more than ±5% from any
integral signal channel number obtained over a period of 8 days.
Within Day
The mean channel value of the integral signal intensity of alignment verification fluorospheres
from the standard PMT’s off the blue red, and violet lasers does not vary more than ±5% from any
integral signal channel number obtained within a period of 24 hours.
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......