773232AH
8-6
Running Samples
Running Samples - MCL Manual Mode
CAUTION
Risk of reporting incorrect results. Data displays for light scatter patterns,
antibody staining profiles, and all gates and boundaries used to arrive at the test
result should be reviewed by a laboratory professional when interpreting the
data. If results are suspect, follow your laboratory procedures to resolve.
10
Print Patient Panel Report, if needed. See
.
Running Samples - MCL Manual Mode
MCL Manual Mode operates the same as MCL Automatic Mode except the MCL stops after processing
each tube to allow you to enter the location of the next tube to process.
The MCL Manual Mode works from a systematic processing of the Worklist in Acquisition Manager
by starting with the sample identified in Row 1, then moving to the sample identified in Row 2, Row
3 etc. Although the tube location may be edited, the order in which sample processing takes place
is always based on the sequential rows in Acquisition Manager. With a partially completed worklist,
the next sample to be processed will always be the next row in the Worklist that has not been run
(completed rows display in blue in Acquisition Manager).
The sequence of processing samples using the MCL Manual Mode is as follows:
•
The Location field in Acquisition Manager auto-fills when the Carousel Number is entered.
•
When MCL Manual Mode is selected, the 'Enter Next Tube Location to process' dialog displays.
•
The 'Enter Next Tube Location to process' dialog’s “Tube Location” field allows a TUBE
LOCATION to be chosen for the selected TEST identified in the row that the arrow is pointing to.
•
The selected test (row identified by the arrow) cannot be altered at this time.
The rules of sample identifiers and listmode file names will always persist when processing samples
regardless of tube location defined. Therefore, prior to manually changing the tube location, ensure
the sample identifiers match the listmode description intended for each specimen.
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......