3-8
Operation Principles
Signal Processing
Integral Signal
Because the total fluorescence in all three cells is the same, but the distribution is different, the
pulse can be integrated to produce an integral signal (see
).
The height of the integral pulse is proportional to the total fluorescence and is obtained when the
cell exits the laser beam. The pulse height, however, represents the most intense amount of
fluorescence produced. The area under the pulse is proportional to the total fluorescence.
Figure 3.8
Integral and Peak Pulses
1.
Direction of sample flow
2.
Cell in laser beam
3.
Peak pulses
4.
Volts
5.
Time
6.
Integral pulses
7.
Volts
8.
Time
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......