773232AH
13-23
Troubleshooting
Cytometer Messages Table
13
System Vacuum Error
During acquisition, the system
vacuum went outside the operating
range and acquisition was stopped.
If error persists, see System Vacuum
Warning.
System Vacuum Warning
Liquid in the vacuum trap.
Check that the vacuum trap (on the front of
the Pneumatic Supply) is tight and is less
than 1/4 full of fluid. If it is more full, empty
it (see
Vacuum line is not connected
between the power module and the
Cytometer.
Connect the vacuum line at the back of the
instrument.
Hardware problem.
contact your local Beckman Coulter
Representative.
TEC Board Not Found
Optics temperature control board
could not be detected.
Restart the entire system. If error persists,
contact your local Beckman Coulter
Representative.
Tube Up/Down Error
Unable to load or unload the sample
tube from the MCL sampling position.
1.
Check that the labels on the sample
tubes are secure and are not adhering to
the walls of the carousel.
2.
Check that there is no crack in the
sample tube.
3.
If problem continues, contact your local
Beckman Coulter Representative
Tube Up/Down Warning
The first attempt to load or unload
the sample tube from the MCL
sampling position failed.
If warning persists, see Tube Up/Down Error.
Violet Laser Comms Warning
Unable to communicate with the
violet laser.
Restart the entire system. If warning
persists, contact your local Beckman
Coulter Representative.
Violet Laser Fault
A fault occurred on the violet laser
during acquisition and the
acquisition was stopped.
Restart the entire system. If the error
persists, contact your local Beckman
Coulter Representative.
Violet Laser Fault Warning
There is a fault condition on the violet
laser.
Check optics covers are properly installed
and tightened. See Violet Laser Fault.
Violet Laser Initialization Error
The violet laser failed to initialize.
Restart the entire system. If the error
persists, contact your local Beckman
Coulter Representative.
Violet Laser Initializing
The violet laser is initializing
If the laser does not initialize within a
reasonable time then restart the entire
system. If the problem persists then contact
your local Beckman Coulter Representative.
Violet Laser Not Calibrated
The instrument is not calibrated to
trigger off the violet laser.
Calibrate the violet laser.
Table 13.2
Cytometer Messages
(Continued)
Message
Probable Cause
Recommended Action
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......