773232AH
10-23
Using Navios Software
Listmode Playback Toolbar
10
Listmode Playback Toolbar
The listmode playback tool utilizes two sets of buttons - one for saving and restoring worklists
(panels populated with listmode files) and one for saving and restoring queues (lists of listmode
files).
Pause / Rotate
Pause Acquisition and rotate sample to the tube access
door.
Stop Acquisition (
(
F10
)
)
Stop the Acquisition of the current sample and output
results.
Abort Acquisition (
(
F12
)
)
Stop the Acquisition without saving data or outputting
reports.
Idle Mode
Place the Cytometer in Idle mode in order to perform
various cleaning and replacing procedures.
Prime
Flush the sample line and flow cell with sheath fluid to
declog or remove bubbles.
Cleanse
Flush the sample line and flow cell with cleaning agent.
MCL Manual Mode
Run a carousel of samples one tube at a time.
Single Tube Mode
Run a single tube in fixed position # 21.
Cytometer Controls
Display the Cytometer Control dialog box to adjust the
Cytometer settings. Click the button again to hide the
dialog box.
Create a new Worklist (clear Worklist)
Save a Panel
Add a Panel to the Worklist
Add a Tube to the Worklist
Delete a tube from the Worklist
Play
Pause
Stop
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......