CHAPTER 3
Operation Principles
What This Chapter Explains
This chapter explains how the Cytometer measures scattered light and fluorescence as cells pass
through the laser beam.
The illustrations in this chapter are not exact representations of the inside of the Cytometer. They
are for explanatory purposes only.
Sample Flow
CAUTION
Possible flow cell damage. To avoid clogging the sample probe, sample tubing or
flow cell, ensure that 12 x 75 mm test tubes are free of debris before you use
them.
Sample Loading
The sample carousel has bar-code labels that identify the carousel and the tube position number.
Also, you can put bar-code labels on the sample tubes. See
The MCL has a bar-code reader that reads the carousel number, the sample tube position, and the
sample tube bar-code labels as the carousel rotates. The MCL handles a sample tube as follows:
•
It lifts the tube out of the carousel into a centering cup.
•
It moves the bottom of the tube in a circular orbit to mix the sample.
•
It lowers its sample probe into the tube and the tube is pressurized. Sample flow begins.
The sample probe is cleaned automatically when sample flow ends.
Hydrodynamic Focusing
The instrument uses a process called hydrodynamic focusing to ensure that the cells move through
the laser beam one at a time, along the same path through the flow cell.
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......