Troubleshooting
Cytometer Messages Table
Optics Temperature Stabilizing
The temperature of the optics is
stabilizing.
If the optics temperature does not stabilize
within a reasonable time then restart the
entire system. If the problem persists then
contact your local Beckman Coulter
Representative.
Optics Temperature Warning
The temperature of the optics differs
considerably from the temperature
when the optics were aligned such
that the instrument may not meet its
specification.
Wait till the temperature of the optics is
closer to that when the instrument was
aligned or realign the optics for different
ambient temperature range.
Overflow Error
The drip or waste container has
overflowed and the acquistion has
been stopped.
Contact your local Beckman Coulter
Representative.
Pico Motor Board Not Found
The pico motor control board could
not be detected.
Restart the entire system. If error persists,
contact your local Beckman Coulter
Representative.
Place tube in position 21
Single Tube Fixed Mode selected
Place tube in position 21 and select Run to
continue in this mode.
To discontinue operation in this mode,
depress the Single Tube Fixed Mode button
on the Cytometer Control toolbar.
Pressures Not Calibrated
Pressures can not be calibrated
because pressure control board can
not be detected or calibration
coefficients corrupted.
Restart the entire system. If error persists,
contact your local Beckman Coulter
Representative.
Probe Up/Down Error
The probe failed to move up or down
correctly after several attempts.
Restart the entire system. If error persists,
contact your local Beckman Coulter
Representative.
Probe Up/Down Warning
The first attempt to raise or lower the
probe failed.
None. The instrument will automatically
retry to raise or lower the probe. If warning
persists, contact your local Beckman
Coulter Representative.
Red Laser Comms Warning
Unable to communicate with red
laser.
Restart the entire system. If the error
persists, contact your local Beckman
Coulter Representative.
Red Laser Fault
A fault occurred on the red laser
during acquisition and the
acquisition was stopped.
Restart the entire system. If the error
persists, contact your local Beckman
Coulter Representative.
Red Laser Fault Warning
There is a fault condition on the red
laser.
Check optics covers are properly installed
and tightened. See Red Laser Fault.
Red Laser Initialization Error
The red laser failed to initialize.
Restart the entire system. If the error
persists, contact your local Beckman
Coulter Representative.
Table 13.2
Cytometer Messages
(Continued)
Message
Probable Cause
Recommended Action
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......