Replace/Adjust Procedures
Replace The Mcl Sample Head
12
10
Discard the old sample head and tubing assembly in accordance with your local regulations and
acceptable laboratory procedures.
11
Place the new tubing manifold into the bracket in the pneumatic drawer and tighten the
thumbscrew.
NOTE
You might find it easier to screw in the thumbscrew with a screwdriver.
12
Route the sample head and tubing through the instrument.