773232AH
12-6
Replace/Adjust Procedures
Replace The 10 L External Sheath Fluid Container
Tighten the cap.
8
Insert the plastic support collar that secures the pickup tube assembly.
9
Place the 10 L external sheath fluid container in a location that is lower than the internal sheath
fluid container. This prevents siphoning of the sheath fluid.
10
Put the cap from the new container onto the old container and dispose of the container
properly.