Troubleshooting
Software Messages
The program will not be able to work
properly because it failed to access
the administrator user settings.
Please close the application
immediately.
The software failed to find the
administrator account.
Reboot the computer, if problem persists
contact your local Beckman Coulter
representative.
The program will not be able to work
properly because it failed to access
the administrator user settings.
Please close the application
immediately.
The software failed to find the
administrator account.
Reboot the computer, if problem persists
contact your local Beckman Coulter
representative.
Unable to open '<session default
protocol name>' in home directory.
\nPlease select a protocol file to copy.
The specified default protocol is not
accessible in the expected location.
Select a protocol file to use as the default
protocol file.
This protocol contains an obsolete
QuadStat region: <region> This
region will not be loaded.
Protocol loaded contains an obsolete
QuadStat
This protocol contains a QuadStat region
from a previous software version and
cannot be converted. Recreate the
QuadStat region
The QC Product file could not be
opened.
The file path is wrong, the name is
wrong or the file is corrupt.
Contact your local Beckman Coulter
representative.
The changes to regions have
resulted in the following regions
having their status as linked regions
removed: <<comma separated list
of region names>>
User has made changes that have
broken the links between regions.
Re-establish the desired links..
The QC Product file could not be
opened.
The QC product file could not be
opened due to the file not being
present in the specified location or
the file is invalid.
Contact your local Beckman Coulter
representative.
Error writing file.
An error occurred writing the file.
Check hard drive space, access rights and
write protection on media.
COleException. SCODE:
<<error-code>>.
An error occurred when opening the
link to Excel.
Close Excel if it is open and retry the export.
If error recurs, save with a filename that has
fewer characters and retry the export.
Error out of memory
System could not allocate memory.
Close down unnecessary applications and
reboot the computer. If problem persists
contact your local Beckman Coulter
representative.
Critical error reading file.
A critical error occurred when
reading the file.
Check for hard drive space and access
rights.
The Instrument Settings file has been
altered. A complete Panel needs to
be run before processing samples.
ASI Panel is aborted.
You must rerun the AutoSetup panel before
acquiring samples.
The file name association is
incomplete or invalid.
The log file is not supported by the
viewing application.
Endure *.log files are associated with the
Notepad application in order to view the log
file.
Table 13.3
Software Messages
(Continued)
Message
Probable Cause
Recommneded Action
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......