Troubleshooting
Cytometer Messages Table
Data Socket Error
A data communication error has
occurred between the workstation
and the instrument.
Check the Ethernet cable between the
workstation and the instrument. Restart the
entire system. If the error persists, contact
your local Beckman Coulter Representative.
Detector X Incorrect Type
The type of detector X does not
agree with the type in the instrument
configuration.
Restart the entire system. If error persists,
contact your local Beckman Coulter
Representative.
Detector X Missing
The instrument is configured for
detector X but no detector could be
found.
Restart the entire system. If error persists,
contact your local Beckman Coulter
Representative.
Drip Chamber Level Error
The drip chamber is full and the
acquisition has been aborted.
Results displayed on the workspace
are incomplete.
To avoid the error in the future, change the
sample concentration or protocol stop
condition to reduce the acquisition time to
be 18 minutes or less.
If the recommended action has been taken
and the error persists, contact your local
Beckman Coulter Representative.
Drip Chamber Level Warning
When this message is displayed
during acquisition there is enough
spare volume for another three
minutes of sample analysis.
Stop the acquisition in a controlled way. To
avoid the message in the future, change the
sample concentration or protocol stop
condition to reduce the acquisition time to
18 minutes or less.
If the recommended action has been taken
and the warning persists, see
Drip Chamber
Level Error
.
Drip Or Waste Overfill Error
The drip or waste chamber has
overflowed and the acquisition has
been stopped.
contact your local Beckman Coulter
Representative.
Event Checksum Failures
The checksum of an event passed
from the instrument to the
workstation is incorrect.
Restart the entire system. If error persists,
contact your local Beckman Coulter
Representative.
Failed to Power Off
The instrument failed to turn off.
Power The Cytometer Only On/off
. If error
persists,contact your local Beckman Coulter
Representative.
Fluidics Board Not Found
The fluidics I/O board could not be
detected.
Restart the entire system. If error persists,
contact your local Beckman Coulter
Representative.
Fluidics I/O Board Voltage Error
The voltages on the fluidics I/O board
are out of range.
Restart the entire system. If error persists,
contact your local Beckman Coulter
Representative.
Front Panel Connection Fault – SCA1
A fault was detected in the
connection between the data
acquisition board SCA 1 and the front
panel display board.
Restart the entire system. If error persists,
contact your local Beckman Coulter
Representative.
Table 13.2
Cytometer Messages
(Continued)
Message
Probable Cause
Recommended Action
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......