773232AH
12-24
Replace/Adjust Procedures
Replace The Sample Probe And Sample Pickup Tubing
8
Lift the sample probe up and out of its holder.
9
Discard the old sample pickup tubing and probe assembly in accordance with your local
regulations and acceptable laboratory procedures.
10
Ensure that the rubber washer
(1 )
and O-ring
(2 )
are positioned correctly on the new sample
probe.
b
c