Troubleshooting
Cytometer Messages Table
Blue Laser Comms Error
Unable to communicate with blue
laser.
Restart the entire system. If error persists,
contact your local Beckman Coulter
Representative.
Blue Laser Fault
A fault occurred on the blue laser
during acquisition and the
acquisition was stopped.
Restart the entire system. If the error
persists, contact your local Beckman
Coulter Representative.
Blue Laser Fault Warning
There is a fault condition on the blue
laser.
Check optics covers are properly installed
and tightened. See Blue Laser Fault.
Blue Laser Initialization Error
The blue laser failed to initialize.
Restart the entire system. If the error
persists, contact your local Beckman
Coulter Representative.
Blue Laser Initializing
The blue laser is initializing
If the laser does not initialize within a
reasonable time then restart the entire
system. If the problem persists then contact
your local Beckman Coulter Representative.
Blue Laser Not Calibrated
The instrument is not calibrated to
trigger off the blue laser.
Calibrate the blue laser.
Blue Laser Power Error
During acquisition the blue laser
output power went out of range and
the acquisition was stopped.
Restart the entire system. If error persists,
contact your local Beckman Coulter
Representative.
Blue Laser Power Warning
The blue laser power is unstable.
If laser does not stabilize within a
reasonable time or frequently goes unstable
then restart the entire system. If problem
persists then contact your local Beckman
Coulter Representative.
CAN Master Board Not Found
The CAN master board could not be
detected
Restart the entire system. If error persists,
contact your local Beckman Coulter
Representative.
Carousel In/Out Error
The carousel failed to move in or out
correctly after several attempts.
1.
Check that there is no obvious
obstruction (sample tube) in the MCL
area.
2.
Check that the 30 psi gauge on the
Pneumatic Supply is okay.
3.
If no obstruction is found and pressure
okay, contact your local Beckman
Coulter Representative.
Carousel In/Out Warning
The first attempt to move the
carousel in or out failed.
If warning persists, see Carousel In/Out
Error.
Carousel Label Read Error
The carousel bar-code label could not
be read after several attempts.
1.
Check that the bar-code label is not torn
or written on.
2.
Try using another carousel.
3.
If problem continues, contact your local
Beckman Coulter Representative.
Table 13.2
Cytometer Messages
(Continued)
Message
Probable Cause
Recommended Action
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......