Analysis Menu
10
Polygonal Region Properties
Region Name
Displays the name of the region. If the name is not grayed, you can rename it.
Prime
When checked, this option allows the system to determine if the peak of the events on the plot is
within this region. If the peak is outside this region (due to an air bubble or partial clog), the system
performs a Prime cycle. A maximum of three sequential Prime cycles are allowed; if the peak of the
plot is still not within this region, the MCL aborts.
Automatic Region Name Positioning
Region labels default to an automatic position. You may move a region label to any desired position
within a plot, this user modification automatically unchecks this option. To revert to the automatic
position recheck this option.
Region Statistics exported for Quality Control
The region's results are exported to the Quality Control feature of the Report Generator.
NOTE
Exporting region statistics and selecting the QC product for FlexQuad regions operates differently
depending on whether the Region Properties dialog is opened via,
•
right click on the selected region in the plot and using the context menu or
•
right click on the region in Protocol explorer or double click the region name in the View /
Modify regions dialog.
If the Region Properties dialog is opened via right click on the selected region in the plot and using
the context menu, when a QC product is selected or changed, it is applied to all quadrants.
Automatic Region Name Positioning
Region statistics exported for QC
NOTE
You cannot make any changes on this screen when
running a
locked protocol other than selecting a
product or choosing to export QC statistics.
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......