773232AH
4-7
Specifications
Software Specifications
4
Parameters
16 different signals, including Time and Ratio (derived parameter) can be acquired simultaneously
in logarithmic and linear modes, giving up to 32 acquisition parameters from a possible 62.
5 different signals available from each detector:
•
Integral linear
•
Integral logarithmic
•
Peak linear
•
Peak logarithmic
•
Time of Flight linear.
Regions
Up to 32 gates and 256 analysis regions are available for gating, analysis, and autogating per
protocol. Up to 8 of those regions can be used to create any one gate. The following types of regions
are available for gating and analysis:
•
Linear
•
Rectangular
•
Quadrant
•
Polygonal
•
Elliptical
•
Contour.
Listmode Analysis
The instrument can store up to 16 parameters including TIME, and RATIO as listmode data. The
amount of available random access memory (RAM) in the computer determines the maximum size
of a listmode file. See
Locked Protocols
User defined protocols can be locked or unlocked by a System Administrator and password
protected to prevent inadvertent modifications. The optional application software
(e.g. Navios tetra) uses locked protocols that cannot be unlocked by a System Administrator . The
following software screens are disabled when using locked protocols. You cannot make changes to
the locked features on these screens.
(QC product selection and export region statistics are not locked)
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......