Using Navios Software
Cytometer Menu
Cytometer Control Acquisition Setup Tab
This screen is used to set the Acquisition Mode and Acquisition Limits. Additionally, the
Discriminator values for the selected Parameters can be set.
See also:
Cytometer Control Compensation Tab
Cytometer Control Parameter Setup
Settings
The Settings grid allows you to specify Label, Discriminator, Voltage and Gain values for each of the
Scatter and Fluorescence detectors installed on the instrument. Detector row headings will be
colored according to the Laser to which the Detector relates. If the Laser or PMT to which a Detector
relates is not installed on the instrument, then all fields on that row will be disabled.
Label
These input fields allow the editing the Names assigned to the default manufacturer’s detectors.
Changes are ONLY reflected in the names shown within the Discriminator box in the Acquisition
Setup tab and the Compensation tab. This allows the various fluorochromes to be easily recognized
when adjusting instrument settings.
NOTE
You cannot make any changes on this
screen when running a
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......