773232AH
4-9
Specifications
Software Specifications
4
The initial protocol in each application, which you use to adjust or confirm correct PMT High
Voltage and parameter gain settings, utilizes Flow-Set Pro fluorospheres.
If you wish to use a negative control sample use the “AS *_STAND.Pro” protocol in the above panels.
You must modify the protocol for the number of colors you are analyzing.
Each AutoSetup II application requires a user defined base protocol with the desired parameter
names to generate the application. The Application Definition Wizard creates a unique base
protocol with the same name as the application.
The guidelines below show approximate region widths generated by the Application Definition
Wizard.
Flow-Set Pro region widths generated by the Application Definition Wizard may be
edited.
If different regions widths are desired, open the appropriate *_Stand protocol onto the workspace
and edit the regions. Save the protocol. Subsequent scheduling of the application uses the edited
regions.
•
Forward Scatter (linear): ±20
•
Side Scatter (linear): ±50
•
Fluorescence Parameters (logarithmic);
•
1st decade - beginning or low end: ±0.1
•
1st decade - high end: ±0.3
•
2nd decade - beginning or low end: ±1.0
•
2nd decade - high end: ±3.0
•
3rd decade - beginning or low end: ±10
•
3rd decade - high end: ±30
•
4th decade - beginning or low end: ±100
•
4th decade - high end: ±300
Default Generic Panels and Protocols
The following generic protocols are included with Navios software.
Application
Name
Flow-Set Pro Protocol
Compensation
Protocols
Verification Protocol
Settings Protocol
Base Protocol
AS 6C 2L
AS 6C 2L_STAND.Pro
AS 6C 2L_FL01.Pro
AS 6C 2L_FL02.Pro
AS 6C 2L_FL03.Pro
AS 6C 2L_FL04.Pro
AS 6C 2L_FL05.Pro
AS 6C 2L_FL06.Pro
AS 6C 2L_VERIFY.Pro
AS 6C 2L
Settings.Pro
AS
6C2L_APP.Pro
Panel Name
Protocol
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......