773232AH
4-16
Specifications
Software Specifications
$PnB
Number of bits allocated to store data for parameter n: 32 = 20 bit data per
parameter plus 12 bits of padding.
$PnE
Lin/Log parameter calibration. This value is always “0,0” and it means linear.
$PnN
Name of parameter n, for example: Manufacturer assigned name, that is: SS-A,
FL2-H, FL3-W.
$PnR
Channel resolution of parameter n: 1048576 = 20 bit parameters.
$SPILLOVER
Compensation matrix. The matrix is size is specified in the first value of the
keyword. The following values indicate the index of each of the parameters listed
in the matrix.
The actual values in the matrix (Sij) are floating point numbers that indicate the
ratio of the amount of signal in the “j” channel to the amount of signal in the “i”
channel for particles carrying only the “i” parameter dye.
Only parameters associated with fluorescence detectors are present in this
compensation matrix.
NOTE
This keyword is not part of the FCS 3.0 standard. It has been included
based on a pre release version of the FCS 3.1; its implementation does not
match the released version of the FCS 3.1 standard.
$TOT
Total number of events acquired.
@BARCODE
Barcode read from the sample loader or “NoRead” if nothing is read.
@BLUELASERPOWER_END
Blue laser power at end of acquisition.
@BLUELASERPOWER_START
Blue laser power at start of acquisition.
@BLUELASERSHUTTER
Blue laser shutter status: “Open” or “Closed”.
@BLUETARGETPOWER
Blue laser target power.
@CHECKSUM
FCS data checksum.
@CAROUSEL
Carousel number read from the sample loader.
@CRS20BITFORMAT
Data stored in CRS 20bit format. This value is always “TRUE”
@CYTOMETERID
The unique ID given to the flow cytometer by the user, for example: the flow
cytometer serial number.
@PnADDRESS
Multiplexer address of parameter n.
@REDLASERPOWER_END
Red laser power at end of acquisition.
@REDLASERPOWER_START
Red laser power at start of acquisition.
@REDLASERSHUTTER
Red laser shutter status: “Open” or “Closed”.
@REDTARGETPOWER
Red laser target power.
@SAMPLEID1
User defined string.
@SAMPLEID2
User defined string.
@SAMPLEID3
User defined string.
@SAMPLEID4
User defined string.
@TESTFILE
Protocol file name.
Table 4.2
FCS 3.0 Keyword Information
FCS 3.0 Keyword
Key Value
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......