Navios Flow Cytometer
Instructions For Use
PN 773232AH (October 2015)
© 2015 Beckman Coulter, Inc.
All rights reserved.
Beckman Coulter, the stylized logo,
COULTER
and
Navios
are
trademarks of Beckman Coulter, Inc. and are registered in
the USPTO.
All other trademarks, service marks, products, or services
are trademarks or registereed trademarks of their
respective holders.
Contact Us
If you have any questions, contact our Customer Support
Center.
•
Worldwide, find us via our website at
www.beckmancoulter.com/customersupport/support.
•
In the USA and Canada, call us at 1-800-369-0333.
•
Outside of the USA and Canada, contact your local
Beckman Coulter Representative.
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......