Specifications
Instrument Specifications And Characteristics
4
Red Laser Beam Spot Size
An elliptical spot 9.6
μ
m high by 72
μ
m wide.
Violet Laser Beam Spot Size
An elliptical spot 8.9
μ
m high by 70
μ
m wide.
Laser Beam Separation
The laser beams are 125
μ
m (±12.5
μ
m) apart.
Optical Filters
The filters used in the Navios system are dependant upon your system configuration. See
,
.
Sensors
•
The FS sensor and the SS sensor are photodiodes.
•
The FL sensors are photo-multiplier tubes (PMTs) that have a 200-nm to 800-nm spectral range.
Signal Processing
•
High voltage amplification, minimum 250 up to 1,100, in increments of 1, for FL1-FL10.
•
Vernier gain (fine amplification), up to 1,000 (labeled volts), in increments of 1, for FS and SS. A
change of 1 to 750 represents a 1-to-4 change in gain:
•
Linear amplification (gain) by 1.0, 2.0, 5.0, 7.5, 10, 20, 50, 75, 100, 200, 500 or 750 for FS and SS.
•
Linear amplification (gain) by 1.0 or 2.0 for FL1-FL10.
•
Four-decade digital logarithmic transformation of FS, SS and FL1-FL10.
Note
: A scale of 0.1 to 1,000 is displayed on the plot axis for logarithmic parameters, but the
default statistics are based on an actual scale of 0.1024 to 1024. The displayed scale can be
changed to 1 to 10,000.
•
Six decade with True View enabled. Data is transformed in a ‘log’ style at high values, but
‘linearly’ near the axis. Up to 2 negative decades of data may be displayed. Transform ‘reflects’
around origin and becomes Logarithmic at high negative values.
•
Fluorescence color compensation is available in 0.1 increments, from 0 to 100%, for FL1-FL10.
•
A discriminator (maximum value of 1,000) is available for any one of the signals. Only one
discriminator can be specified for any one sample acquisition.
Dynamic range
20-bit data acquisition
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......