Replace/Adjust Procedures
Replace An Optical Filter
Identify Coated Side Of Dichroic Filter
to determine correct orientation.
5
Place the metal ring over the filter in the filter holder.
6
For the metal ring fastener, insert the special tool into the metal ring’s two slots and turn to the
right to tighten.
NOTE
You might find it easier to begin tightening the metal ring by turning it with your gloved fingers.