Using Navios Software
Analysis Menu
10
Minimum Count
Check this option and enter a number in the
Count
field to enable the minimum count function. This
overrides an existing stop count to ensure a minimum number of events within a particular region
are collected. The actual stop count in this case is approximate (but is always in excess of the
minimum count value). This is limited by the Duration and Total Events.
Linked Region
Select a region to be linked to another region and the region tracks the selected region. A region
may only link to a region of the same type and parameters. For dual parameter regions, the
parameters may be transposed. A polygonal region may link to a quadrant region/FlexQuad region,
but a quadrant region/FlexQuad region may not link to a polygonal region.
When a region is linked to another, changes to either region's coordinates (by dragging the region)
are propagated to both regions. The exception is when the region is an auto analysis region
(positives analysis, Elliptical and Contour AutoGate). Then the regions track the auto analysis region
and the link is unidirectional.
If a region is linked to another, positives analysis and autogating are disabled for that region.
Polygonal Region Format Tab
Interactive Polygonal Region Editing
When a Region's position is changed, any Gates and statistics that use the Region are automatically
updated and each plot is redrawn to reflect the changes in that Region. When the Region is deleted,
any Gate or statistics that used the Region are also automatically deleted. When a Gate has been
deleted, any plot that used the deleted gate is reset to Ungated. Navios software allows you to resize
an already created Region.
Rectangular Region
This option draws a Rectangular Region on a dual parameter plot. See also:
.
Specify the
Color
and line
Weight
for the Polygonal
Region boundary.
NOTE
You cannot make any changes on this screen when
running a
.
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......