773232AH
12-37
Replace/Adjust Procedures
Adjust The System Pressure
12
1
Open the Supply Cart front door and locate the Pressure Adjustment knob (1).
2
Pull the collar around the Pressure Adjust knob out toward you.
3
Adjust the pressure to 30 ±2 psi.
(1) To decrease, turn to the left.
b
PU
LL
TO ADJU
ST