Using Navios Software
Analysis Menu
If the Region Properties dialog for any of the quadrants is opened via right click in Protocol explorer
or double clicking the name in the View / Modify regions dialog, then changing a QC product for a
quadrant is only applied to this quadrant.
QC Product
Select the QC product from the drop down list. See also,
.
NOTE
In a networked database configuration, when QC Product is added or changed by one user it is not
visible to another user in the networked configuration unless the user logs out and logs back in to refresh
the Region Properties dialog box QC Product list.
AutoGating
None
This is the default when the AutoGate option is not selected.
Elliptical
Toggle the AutoGate to an elliptical region.
Contour
Toggle the AutoGate region to a contour region.
Sensitivity
Eliptical AutoGate settings are 0.11, 0.33 and 1.1 (% of processed peak). Contour AutoGate settings
are 1, 2, 3, 4, and 5.
•
Level 1 = 0.11% processed peak at a resolution of 64x64.
•
Levels 2, 3 and 4 = 0.11, 0.33 and 1.1% at a resolution of 128x128.
•
Level 5 = 1.1 % processed peak at a resolution of 256x256.
Selecting 0.11 creates a larger region, whereas selecting 1.1 creates a smaller radius region around
the target population.
Travel
Input a maximum value the contour AutoGate region travels from the mean. The Minimum value is
25; Maximum is 200 channels. Use this option to slightly adjust the AutoGate from tube to tube if
slight changes in lysis cause the target population to move slightly. This option is not available for
elliptical AutoGates.
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......