773232AH
3-2
Operation Principles
Laser Beam Shaping
The flow cell (
) contains a rectangular channel. A pressurized stream of sheath fluid enters
the channel at the lower end and flows upward. The sensing area of the flow cell is at the center of
the channel.
While the sheath stream is flowing through the channel, a stream of sample is injected into the
middle of the sheath stream. As shown in
, the sheath stream surrounds, but does not mix
with, the sample stream. The pressure of the sheath stream focuses the sample stream so that the
cells flow through the laser beam single file. If the cells were to move through the laser beam in
different ways during sample flow, sample analysis could be distorted.
Figure 3.1
Flow Cell
Laser Beam Shaping
Before the laser beams reach the sample stream, cross-cylindrical lenses focus the beams (see
). Focusing keeps the beam perpendicular to the sample stream flow while making the
beam small enough to illuminate only one cell at a time.
The first lens controls the width of the beam; the second, the height. The resulting elliptical beam
is focused on the sensing area of the flow cell.
1.
Waste out
2.
Debubble port
3.
Sheath stream enters here
4.
Sample stream enters here
b
c
d
e
Summary of Contents for Navios
Page 8: ...773232AH viii Safety Notice Safety Symbols ...
Page 34: ...xxxiv Tables A 2 Code Related Specifications A 6 ...
Page 38: ...773232AH xxxviii Introduction Graphics ...
Page 44: ...773232AH 1 6 Use And Function Material Safety Data Sheets MSDS ...
Page 72: ...773232AH 3 18 Operation Principles Plot Display ...
Page 90: ...773232AH 4 18 Specifications Software Specifications 1 2 3 ...
Page 140: ...773232AH 5 34 System Overview Using The System Help ...
Page 154: ...773232AH 6 14 Daily Routine Cytometer Auto Startup ...
Page 194: ...773232AH 8 16 Running Samples Data Review ...
Page 412: ...773232AH 10 216 Using Navios Software Report Generator Error Log ...
Page 446: ...773232AH 11 34 Cleaning Procedures Power The Cytometer Only On off ...
Page 522: ...773232AH 13 30 Troubleshooting Cytometer Control Window Cannot Be Viewed ...
Page 530: ...773232AH A 8 Bar code Specifications Checksum Algorithm ...
Page 540: ...773232AH References 2 References ...
Page 582: ...Index 36 Index ...
Page 584: ...773232AH Warranty 2 Beckman Coulter Inc Customer End User License Agreement ...
Page 585: ......