Appendix D Performance Test Record Form
D-4
Table D.1-3 Pulse Pattern Generator Performance Test Record Form (With MP2110A-093)
Frequency Accuracy
Connector
Set Bit Rate
(kbit/s)
Minimum
Value (kHz)
Measured
Value (kHz)
Maximum
Value (kHz)
Pass/Fail
Sync Out
28200000
3524964.8
3525035.2 Pass/Fail
9500000
1187488.2
1187511.8 Pass/Fail
Measurement Channel PPG 1 2 3 4
Waveform (Amplitude)
Connector Set Amplitude
Minimum
Value
Measured
Value
Maximum
Value
Pass/Fail
Data Out
0.1 Vp-p
0.06 Vp-p
Vp-p
0.14 Vp-p Pass/Fail
0.8 Vp-p
0.62 Vp-p
Vp-p
0.98 Vp-p Pass/Fail
Out
0.1 Vp-p
0.06 Vp-p
Vp-p
0.14 Vp-p Pass/Fail
0.8 Vp-p
0.62 Vp-p
Vp-p
0.98 Vp-p Pass/Fail
Waveform
Connector
Item
Minimum
Value
Measured
Value
Maximum
Value
Pass/Fail
Data Out
Crossing
40%
%
60% Pass/Fail
Rise Time
ps
17 ps Pass/Fail
Fall Time
ps
17 ps Pass/Fail
Jitter (rms)
ps
0.9 ps Pass/Fail
Out
Crossing
40%
%
60% Pass/Fail
Rise Time
ps
17 ps Pass/Fail
Fall Time
ps
17 ps Pass/Fail
Jitter (rms)
ps
0.9 ps Pass/Fail
Skew
Connector Set Amplitude
Minimum
Value
Measured
Value
Maximum
Value
Pass/Fail
Data Out,
Out
0.3 Vp-p
ps
±8 ps Pass/Fail
Data
Data
Data
Summary of Contents for BERTWave Series
Page 26: ...VI...
Page 74: ...Chapter 1 Outline 1 48...
Page 166: ...Chapter 4 Screen Operation 4 24...
Page 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Page 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Page 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Page 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Page 432: ...Appendix A Specifications A 36...
Page 458: ...Appendix D Performance Test Record Form D 12...
Page 466: ...Index Index 6...