Chapter 5 How to Operate BERT
5-20
Threshold Level
Voltage level for evaluating “1” and “0”.
The input connectors are AC coupled, so set a voltage corresponding to the
signal waveform excluding the DC components.
The waveform excepting the waveform input to the connector and the DC
component when inputting the 1.2V LVCMOS signal to the
Data In
connector and
Data
In
connector is shown below.
Set the threshold level to the waveform excepting the DC component.
Figure 5.2.8-2 Waveform That Threshold Level Can be Set
Logic
Select the positive logic (POS) or negative logic (NEG).
Voltage (V)
Waveform without DC component
Time
Voltage (V)
Waveform input to
Data In
–
Data
തതതതതതത
In
connectors
Time
0
0
Threshold level
1
1
–
1
–
1
Summary of Contents for BERTWave Series
Page 26: ...VI...
Page 74: ...Chapter 1 Outline 1 48...
Page 166: ...Chapter 4 Screen Operation 4 24...
Page 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Page 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Page 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Page 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Page 432: ...Appendix A Specifications A 36...
Page 458: ...Appendix D Performance Test Record Form D 12...
Page 466: ...Index Index 6...