
Chapter 7 Performance Test
7-4
7. Read the value measured by the frequency counter.
8. Check that the read value is within the following limited values.
Without the MP2110A-093: 3 037 500±30.37 kHz
With the MP2110A-093:
1 187 500±11.87 kHz
7.1.3
Waveform
(1) Specification
Amplitude:
0.1 to 0.8 Vp-p
Accuracy:
(±20% of setting) ±20 mV
Data Crossing:
50±10%
(Amplitude: 0.3 Vp-p, Bit rate: 25.78125 Gbit/s)
Rise/fall time:
17 ps
(Amplitude: 0.3 Vp-p, Percentage: 20-80%, Bit
rate: 25.78125 Gbit/s)
Jitter (RMS):
0.9 ps (Amplitude 0.3 Vp-p, 25.78125 Gbit/s)
(2) Setup
Figure 7.1.3-1 and Figure 7.1.3-2 show the connection diagrams
when performing the test using one MP2110A.
Figure 7.1.3-3 and Figure 7.1.3-4 show the connection diagrams
when performing the test using two MP2110As (one MP2110A is used
as the sampling oscilloscope).
Figure 7.1.3-1 PPG1 Data Out Waveform Test Connection Diagram
Coaxial Terminator
Summary of Contents for BERTWave Series
Page 26: ...VI...
Page 74: ...Chapter 1 Outline 1 48...
Page 166: ...Chapter 4 Screen Operation 4 24...
Page 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Page 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Page 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Page 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Page 432: ...Appendix A Specifications A 36...
Page 458: ...Appendix D Performance Test Record Form D 12...
Page 466: ...Index Index 6...