7.2 Performance Test for Error Detector
7-17
7
Performance T
est
7.2.3
Rx sensitivity
(1) Specification
40 mVp-p
BER 10
–12
or less under the following conditions:
Bit rate:
25.78125 Gbit/s
Pattern:
PRBS31
Mark
ratio: 50%
Single-end
Back-to-back
connection
(2) Setup
Same as Figure 7.2.2-1 and Figure 7.2.2-2.
(3) Procedure
1. Connect
the
Clk Out
connector and the
Trigger Clk In
connector
of the sampling oscilloscope using a coaxial cable.
2. Connect the coaxial terminators to the
PPG1
Data
Out
and
ED1
Data
In
connectors.
3. Attach a 20 dB fixed attenuator to the
PPG1 Data Out
connector.
(Refer to (a) in Figure 7.2.2-1.)
4. Connect the 20 dB fixed attenuator and
Ch A In
connector using
a coaxial cable.
5. Click
PPG/ED Ch1. Configure the settings as shown below:
Item
Setting Value
Reference Clock
Internal
Symbol Rate
100GbE/4 (25.78125G), 0 ppm
Clock Output
Ch1/2
PPG Amplitude
0.4
External ATT
0
Test Pattern (PPG)
PRBS 2^31–1, POS
Test Pattern (ED)
PRBS 2^31–1, POS
ED Input Condition Single-Ended Data
Threshold 0
mV
PPG Data/XData
ON
Gating Cycle
Single
Gating Period
45 s
Summary of Contents for BERTWave Series
Page 26: ...VI...
Page 74: ...Chapter 1 Outline 1 48...
Page 166: ...Chapter 4 Screen Operation 4 24...
Page 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Page 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Page 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Page 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Page 432: ...Appendix A Specifications A 36...
Page 458: ...Appendix D Performance Test Record Form D 12...
Page 466: ...Index Index 6...