Chapter 6 How to Operate Sampling Scope
6-54
Table 6.2.6-1 Items of Time Dialog Box
Tab
Item
Description
Rat
e
Data Clock Rate
Sets input data speed, clock frequency, and frequency
dividing rate.
Tracking*
1
Specifies Symbol Rate whether to follow the Bit Rate of
PPG/ED Ch1.
Off:
Symbol Rate does not follow PPG Bit Rate nor CRU
Rate.
Symbol Rate: PPG, Divide Ratio: Clock Output:
Symbol Rate follows PPG Bit Rate.
Divide Ratio follows PPG Clock Out.
Symbol Rate: PPG, Divide Ratio: Sync Output:
Symbol Rate follows PPG Bit Rate.
Divide Ratio follows PPG Sync Out.
Symbol Rate: PPG, Divide Ratio: User Defined:
Symbol Rate follows PPG Bit Rate.
Symbol Rate: CRU: 2*
2
:
Symbol Rate follows CRU Rate. Divide Ratio is
determined as follows, depending on which CRU to
track and its setting.
MP2110A-054: 2
MP2110A-055: 4 when the operation rate is 20G
band.
8 when the operation rate is 50G
band.
Tracking Target*
3
Sets which CRU to track the setting when Symbol Rate: CRU
is selected for Tracking.
26G:
26G CRU
53G:
53G CRU
Recalculate Option
Clock Rate:
Calculates Clock Rate from Symbol Rate and
Divide Rate.
Symbol Rate: Calculates Symbol Rate from Clock Rate and
Divide Rate.
Clock Rate
Sets clock frequency input to
Trigger Clk In
connector.
Divide Ratio
Sets frequency dividing rate (Symbol Rate/Clock Rate).
Symbol Rate
When Tracking is Off, Sets signal symbol rate to be
measured.
Acquire Clock Rate
Measures clock frequency input to
Trigger Clk In
connector.
Divide Ratio Detect
Sets whether to detect frequency dividing ratio of clock signal
input to
Trigger Clk In
connector. The automatic detection of
the dividing ratio and the Divide Ratio resetting are
preformed when this item is set to On or Auto Scale is
executed.
Precision Trigger*
4
On, Off
Enables or disables the precision trigger.
Reset
Resynchronizes the trigger when the precision trigger is set
to On.
*1: An error message is displayed because the Tracking operation is not
performed when one or both of the following conditions is/are met.
Summary of Contents for BERTWave Series
Page 26: ...VI...
Page 74: ...Chapter 1 Outline 1 48...
Page 166: ...Chapter 4 Screen Operation 4 24...
Page 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Page 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Page 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Page 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Page 432: ...Appendix A Specifications A 36...
Page 458: ...Appendix D Performance Test Record Form D 12...
Page 466: ...Index Index 6...