6.2 Explanation
of
Windows
6-13
6
How to Operate Sampling Scope
Status Display
The following statuses of the sampling scope are displayed.
Table 6.2.1-1 Status Display Items
Item
Color
Description
CAL
Orange The correction value of O/E is not
appropriate. Calibrate the O/E module
according to 6.3.2, “Adjusting Dark Current”.
CAL
Yellow There is a difference of ± 2.5 °C or more from
the calibration temperature.
Calibrate the level according to 6.3.1,
CAL
Red
There is a difference of ± 5.0 °C or more from
the calibration temperature.
Calibrate the level according to 6.3.1,
26G CRU Unlock Orange MP2110A-054 Clock Recovery
(Electrical/Optical) is not locked on the input
signal.
53G CRU Unlock Orange MP2110A-055 26G/53Gbaud Clock Recovery
(SM Optical) is not locked on the input
signal.
Free Running
Red
The trigger clock cannot be detected.
PT phase unlock
*
Orange Precision Trigger is not locked on the input
signal.
Trigger setting
wrong
Orange The trigger clock can be detected, but its
frequency does not match the clock frequency
setting value
*: Displayed when the MP2110A-024 is installed.
Summary of Contents for BERTWave Series
Page 26: ...VI...
Page 74: ...Chapter 1 Outline 1 48...
Page 166: ...Chapter 4 Screen Operation 4 24...
Page 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Page 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Page 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Page 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Page 432: ...Appendix A Specifications A 36...
Page 458: ...Appendix D Performance Test Record Form D 12...
Page 466: ...Index Index 6...