Chapter 7 Performance Test
7-10
4. Click
PPG/ED Ch1. Configure the settings as shown below:
Item
Setting Value
Reference Clock
Internal
Bit Rate
100GbE/4 (25.78125G), 0 ppm
Clock Output
Ch1/2
PPG Amplitude
0.3 Vp-p
External ATT
0 dB
Test Pattern (PPG)
PRBS 2^9–1, POS
PPG Data/XData
ON
5. Click
Scope. Configure the settings as shown below:
Dialog Box
Item
Setting Value
Setup Signal Type
NRZ
Sampling
Mode
Eye
Number
of
Samples
4050
Accumulation
Time
Persistency
Time
10.0
sec
6. Display the X1 marker by clicking Marker of Scope, and move
the X1 marker to the left cross-point.
7. Connect the coaxial terminator to the
PPG1 Data Out
connector.
8. Connect
the
PPG1
Data
Out
connector to the
Ch A In
connector
of the sampling oscilloscope. (refer to Figure 7.1.4-2).
9. Display the X2 marker by clicking Marker of Scope, and move
the X2 marker to the left cross-point.
10. When the time unit is UI, change Unit in the Scale/Offset tab to
Time by clicking Time of Scope.
11. Measure the time lag of two waveforms using the X1 and X2
markers. Note that the polarities of two waveforms are inverted.
Summary of Contents for BERTWave Series
Page 26: ...VI...
Page 74: ...Chapter 1 Outline 1 48...
Page 166: ...Chapter 4 Screen Operation 4 24...
Page 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Page 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Page 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Page 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Page 432: ...Appendix A Specifications A 36...
Page 458: ...Appendix D Performance Test Record Form D 12...
Page 466: ...Index Index 6...