Appendix D Performance Test Record Form
D-2
D.1 Pulse Pattern Generator
Measurement Uncertainty
The measurement uncertainly is up to double the amplitude
measurement accuracy of the measuring instrument. When a value of
which the measurement uncertainly is added or subtracted to/from the
measurement value is in the range of the specifications, the value passes.
The measurement uncertainty of the frequency accuracy when using
MF2412C Microwave Frequency Counter as the frequency counter is
shown below.
Without MF2412C-003:
0.1 ppm
With MF2412C-003:
0.01 ppm
Refer to Amplitude accuracy in Table A.3.9-2 “Electrical Channel
Performance” for the amplitude measurement uncertainty when using
MP2110A as the sampling oscilloscope. Examples of calculating the
amplitude measurement uncertainty are listed below.
Table D.1-1 Amplitude Measurement Uncertainty
Measured
Value (V)
Measurement
Accuracy (mV)*
Measurement
Uncertainty (mV)
0.96
19.6
39.2
0.80
17.2
34.4
0.64
14.8
29.6
0.12
7.0
14.0
0.10
6.7
13.4
0.08
5.5
11.0
*: The measurement uncertainty when Scale and Offset are set as
follows:
Scale:
100 mV/div
Offset: 0 mV
Summary of Contents for BERTWave Series
Page 26: ...VI...
Page 74: ...Chapter 1 Outline 1 48...
Page 166: ...Chapter 4 Screen Operation 4 24...
Page 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Page 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Page 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Page 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Page 432: ...Appendix A Specifications A 36...
Page 458: ...Appendix D Performance Test Record Form D 12...
Page 466: ...Index Index 6...