Chapter 7 Performance Test
7-20
3. Click
PPG/ED Ch1. Configure the settings as shown below:
Item
Setting Value
Reference Clock
Internal
Symbol Rate
100GbE/4 (25.78125G), 0 ppm
Clock Output
Ch1/2
PPG Amplitude
0.8
External ATT
0
Test Pattern (PPG)
PRBS 2^31–1, POS
Test Pattern (ED)
PRBS 2^31–1, POS
ED Input Condition Single-Ended Data
Threshold 0
mV
PPG Data/XData
ON
Gating Cycle
Single
Gating Period
45 s (28.2 Gbit/s or 24.3 Gbit/s)
120 s (9.5 Gbit/s)
4. Click
Start at All Measurements.
5. After measurement, confirm that the error count of the ED
Result is 0.
6. Change Test Pattern for (PPG) and (ED) in succession to PRBS
2^7–1, PRBS 2^9–1, PRBS 2^15–1, and PRBS 2^23–1, and
repeat steps 4 and 5 each time.
7. Connect coaxial terminators to the
PPG1 Data Out
and
ED1
Data In
connectors. (Refer to Figure 7.2.4-2.)
8. Connect
the
PPG1
Data
Out
and
ED1
Data
In
connectors using
a coaxial cable.
9. Click
PPG/ED Ch1 and set ED Input condition to Single-Ended
XData.
10. Repeat steps 3 to 6.
11. Change the Bitrate to the following value,
When MP2110A-093 is not installed:
24.3 Gbit/s
When MP2110A-093 is installed:
9.5 Gbit/s
12. Repeat steps 1 to 10.
For ED2 to ED4, perform tests in the same way. For ED3 and ED4 test,
change the Clock Output in step 3 to the following value.
Item
Setting Value
Clock Output
Ch3/4
Summary of Contents for BERTWave Series
Page 26: ...VI...
Page 74: ...Chapter 1 Outline 1 48...
Page 166: ...Chapter 4 Screen Operation 4 24...
Page 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Page 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Page 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Page 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Page 432: ...Appendix A Specifications A 36...
Page 458: ...Appendix D Performance Test Record Form D 12...
Page 466: ...Index Index 6...