7.3 Performance Test for Sampling Oscilloscope
7-29
7
Performance T
est
With the MP2110A-021, test the amplitude accuracy for Ch B after
measuring the amplitude accuracy for Ch A.
1. Click
Scope. Configure the settings as shown below:
Dialog Box
Item
Setting Value
Amplitude
Ch B Scale
50 mV/Div.
Ch B Offset
50 mV
Ch B Attenuation
0 dB
Histogram Target
Channel
Channel
B
2. Set the DC voltage to 0 V.
3. Connect the DC Power Source to
Ch B In
connector.
Connect the center conductor of the coaxial cable to plus and the
shield to minus.
4. Click
Ch B to set to On.
5. Click
Sampling to set to Run.
6. Record a standard deviation (Std Dev) and an average value
(Mean) of the histogram measurement result.
7. Set the DC voltage to 200 mV.
8. Record the average value of the histogram measurement result.
9. Set the DC voltage to –200 mV.
10. Record the average voltage of the histogram measurement
result.