7.2 Performance Test for Error Detector
7-13
7
Performance T
est
7.2.2
Operating frequency
(1) Specification
Without MP2110A-093:
24.3 Gbit/s –100 ppm to 28.2 Gbit/s +100 ppm
With MP2110A-093:
24.3 Gbit/s –100 ppm to 28.2 Gbit/s +100 ppm,
9.5 Gbit/s –100 ppm to 14.2 Gbit/s +100 ppm
BER 10
–12
or less under the following conditions:
Amplitude: 0.05
Vp-p
Pattern:
PRBS2^31–1
Mark
ratio: 50%
Single-end
Back-to-back
connection
(2) Setup
Figure 7.2.2-1 ED1 Data In Input Sensitivity Test Connection Diagram
Coaxial Terminator
Fixed Attenuator
Coaxial Terminator
Fixed Attenuator
(a) Amplitude Check
(b) Input Level Measurement