6.2 Explanation
of
Windows
6-39
6
How to Operate Sampling Scope
Jitter Measure Dialog Box
Clicking Jitter at Measure Setup displays the Jitter Measure dialog box.
If Sampling in the Setup dialog box is Advanced Jitter, the Algorithm tab
and Advanced tab are displayed in the Jitter Measure dialog box.
Figure 6.2.4-7 Jitter Measure Dialog Box - Algorithm Tab
Table 6.2.4-4 Items of Jitter Measure Dialog Box - Algorithm Tab
Item
Description
PDJ Measurement
Sets execution of PDJ measurement. PDJ (Pattern Dependent Jitter) is
jitter measured by applying a band-pass filter to DDJ.
ON: Displays PDJ vs Bit graph.
OFF: Displays DDJ vs Bit graph.
Standard
Specify the standard used for PDJ measurement from the following.
STM-0, STM-1, STM-4, STM-16, STM-64, STM-256
PDJ Filter
Set the combination of filters to be used for PDJ measurement from the
following:
LP, HP0+LP, HP1+LP, HP1'+LP, HP2+LP, HP+LP, HP'+LP, LP',
HP0+LP'
Table 6.2.4-5 shows the filter name and frequency range.
Measurement Edge
Type
Set edge detection method of pattern data from All, Falling or Rising.
The name you set will be displayed in the graph.
Summary of Contents for BERTWave Series
Page 26: ...VI...
Page 74: ...Chapter 1 Outline 1 48...
Page 166: ...Chapter 4 Screen Operation 4 24...
Page 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Page 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Page 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Page 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Page 432: ...Appendix A Specifications A 36...
Page 458: ...Appendix D Performance Test Record Form D 12...
Page 466: ...Index Index 6...