Chapter 6 How to Operate Sampling Scope
6-102
Clicking Center moves the marker without depending on the waveforms.
In this case, the setting values of X1,
Δ
X, Y1, and
Δ
Y are initialized.
Although the marker is not displayed at Display Off, the marker position
can be adjusted by changing the value in the text box.
To limit mask area
To examine in which part of the mask at the center the error occurs when
the error occurs in the mask test, the mask area can be limited.
Setting width and angle can limit the mask area.
If the mask area is limited, the error that occurs in an upper and lower
mask area is not measured.
Figure 6.9.2-4 Example of Limiting Mask Area
1. Click
Measure.
2. Click
Mask Test tab
3
Click the Target Channel button to set the channel.
4. Click the Mask Area Restriction button to set On.
5. Click
the
Angle text box to set the angle within the range of –90 to
90.
6. Click
the
Width text box to set the width within the range of 0.01 to
1.00.
Top
Bottom
Center
Angle 0
˚
Width 0.1 UI
Angle 45
˚
Width 0.1 UI
Angle 45˚
Width 0.3 UI
0.1 UI
0.3 UI
Summary of Contents for BERTWave Series
Page 26: ...VI...
Page 74: ...Chapter 1 Outline 1 48...
Page 166: ...Chapter 4 Screen Operation 4 24...
Page 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Page 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Page 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Page 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Page 432: ...Appendix A Specifications A 36...
Page 458: ...Appendix D Performance Test Record Form D 12...
Page 466: ...Index Index 6...