Appendix A Specifications
A-30
Table A.3.10-2 Optical Channel Performance (Cont’d)
Item
Specifications
Optical Power Meter
Range
When MP2110A-022, 023, 025, 026, 032, 033, 035, 036, 039, 042, 043, 045,
046, or 049 is installed:
–18 to 0 dBm (SMF Input)
–18 to +2 dBm (MMF Input)
When MP2110A-030 or 040 is installed:
–18 to +2 dBm (SMF Input)
Accuracy
Typical ±0.55 dB
(0 to +2 dBm)
Typical ±0.35 dB
(–12 to 0 dBm)
Typical ±0.6 dB
(–18 to –12 dBm)
OE Monitor Out
Signal terminal to output after O/E conversion for CRU connection, added
when MP2110A-054 is installed.
Conversion Gain
When MP2110A-030 is installed:
<SMF Input>
Typical 48 V/W
When MP2110A-022, 023, 025, 026, 032, 033, 035, 036, or 039 is installed:
<SMF Input>
Typical 60 V/W
<MMF Input>
Typical 33 V/W
When MP2110A-040 is installed:
<SMF Input>
Typical 42 V/W
When MP2110A-042, 043, 045, 046, or 049 is installed:
<SMF Input>
Typical 53 V/W
<MMF Input>
Typical 29 V/W
Connector
K Connector (f)
Summary of Contents for BERTWave Series
Page 26: ...VI...
Page 74: ...Chapter 1 Outline 1 48...
Page 166: ...Chapter 4 Screen Operation 4 24...
Page 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Page 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Page 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Page 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Page 432: ...Appendix A Specifications A 36...
Page 458: ...Appendix D Performance Test Record Form D 12...
Page 466: ...Index Index 6...