Chapter 7 Performance Test
7-28
3. Click
Scope. Configure the settings as shown below:
Dialog Box
Item
Setting Value
Setup Signal Type
NRZ
Sampling
Mode
Eye
Number
of
Samples
2048
Accumulation
Time
None
Time Tracking
Off
Divide
Ratio
8
Unit
UI
UI On Screen
2
Software
Delay
0
Amplitude
Ch A Scale
50 mV/Div.
Ch A Offset
50 mV
Ch A Attenuation
0 dB
Histogram Axis
Amplitude
X1
0
X2
2
Y1
–250
mV
Y2
250
mV
4. Set the Current Limit of DC Power Source to 20 mA.
5. Set the DC voltage to 0 V.
6. Connect the DC Power Source to
Ch A In
connector.
Connect the center conductor of the coaxial cable to plus
terminal and the shield to minus terminal.
7. Click
Ch A to set to On.
8. Click
Sampling to set to Run.
9. Record a standard deviation (Std Dev) and an average value
(Mean) of the histogram measurement result.
10. Set the DC voltage to 200 mV.
11. Record the average value of the histogram measurement result.
12. Set the DC voltage to –200 mV.
13. Record the average value of the histogram measurement result.
Summary of Contents for BERTWave Series
Page 26: ...VI...
Page 74: ...Chapter 1 Outline 1 48...
Page 166: ...Chapter 4 Screen Operation 4 24...
Page 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Page 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Page 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Page 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Page 432: ...Appendix A Specifications A 36...
Page 458: ...Appendix D Performance Test Record Form D 12...
Page 466: ...Index Index 6...