Chapter 6 How to Operate Sampling Scope
6-58
Calibration and Adjustment
6.3
6.3.1
Calibrating Level
The amplitude accuracy of the sampling oscilloscope is guaranteed after
performing calibration.
When the calibration is required, the error message “Calibration is
required” is displayed in the Setup dialog box in red.
Perform the calibration, when using the MP2110A for the first time or the
message is displayed.
The calibration procedures are as follows:
1. Check that no signals are input to the front-panel
Ch A In
,
Ch B In
, or
Trigger Clk In
connectors. If there are
Ch C
and
Ch D
, check that no
signals are input to these connectors.
2. Click
Setup to open the Setup dialog box.
3. Click
the
Utilities tab.
4. Click
Calibration, and then the message is displayed to confirm that
the signal is not input to the connector.
5. Click
OK.