Chapter 6 How to Operate Sampling Scope
6-68
6.5.2
Setting Clock Rate and Divide Ratio
The symbol rate can be set by measuring the clock rate from the signal
input to the
Trigger Clk In
connector of the front panel.
1. Click
Time.
2. Click
the
Rate tab.
3. Click the Tracking button of the Data Clock Rate to set to Off.
4. Click the Recalculate Option button to select Symbol Rate.
5. Click
Acquire Clock Rate.
The frequency is displayed in the Clock Rate text box.
Confirm the waveform and signal level input to the
Trigger Clk In
connector when the frequency is not displayed.
6. When the clock frequency is not displayed at Clock Rate or the
displayed frequency is not correct, click the Clock Rate text box and
input the frequency.
7. Click the Divide Ratio text box and input the divide rate.
The symbol rate is calculated by the divide rate and clock rate.
When using the BERT Sync Out as an external clock, the divide ratio is
set as follows.
Sync Output settings
Pattern
Divide Ratio
1/8 Clock
–
8
1/16 Clock
–
16
1/40 Clock
–
40
Pattern Sync
PRBS 2^7–1
127
PRBS
2^9–1
511
PRBS
2^15–1
32767
PRBS
2^23–1
8388607
PRBS
2^31–1
214748647
When setting Sync Output to Pattern Sync, the data cannot be collected.
Summary of Contents for BERTWave Series
Page 26: ...VI...
Page 74: ...Chapter 1 Outline 1 48...
Page 166: ...Chapter 4 Screen Operation 4 24...
Page 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Page 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Page 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Page 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Page 432: ...Appendix A Specifications A 36...
Page 458: ...Appendix D Performance Test Record Form D 12...
Page 466: ...Index Index 6...