Appendix A Specifications
A-34
Table A.3.12-3 Recovered Clock Output
Item
Specifications
Output Mode
The signal output from CRU Output can be selected from Recovered Clock or
input signal.
The through output of the input signal can be output only when the input
signal is Clock.
Recovered Mode
When Recovered Clock Out is set to Recovery Clock
Amplitude
Typical
440 mVp-p
Min.
340 mVp-p
Clock Frequency
6.375 to 7.25 GHz (1/8 Clock for 53 Gbaud Operation and 1/4 Clock for 26
Gbaud Operation)
Additive Jitter
(RMS)
Typical
200 fs*
1,
*
2,
*
3
Max. 400 fs*
1,
*
2
Loop Bandwidth
Selectable from the list as follows.
4MHz
10MHz
Bitrate/1667
Attenuation: –20 dB/dec
Lock Detection
Detection function is available.
Through Mode
When Recovered Clock Out is set to Through
Amplitude
Typical
220 mVp-p
Min. 200 mVp-p
Operation
Frequency
0.1 to 1.81 GHz, 3.19 to 3.625 GHz (3.625 GHz is the 1/16 Clock of 58 GHz)
Additive Jitter
(RMS)
Typical
200 fs*
1,
*
3
Max.
400 fs*
1
Connector
SMA, Connector (f)
Termination
50
Ω
, AC coupled
Waveform Format
Square wave
*1: 26.5625 Gbaud, 53.125 Gbaud, when inputting a clock pattern, Outer
OMA = 0 dBm input
*2: LBW = 4 MHz
*3: At 25±5°C
Summary of Contents for BERTWave Series
Page 26: ...VI...
Page 74: ...Chapter 1 Outline 1 48...
Page 166: ...Chapter 4 Screen Operation 4 24...
Page 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Page 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Page 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Page 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Page 432: ...Appendix A Specifications A 36...
Page 458: ...Appendix D Performance Test Record Form D 12...
Page 466: ...Index Index 6...