Chapter 3 Measurement Examples
3-4
3.2 Measuring Waveform
When measuring the DUT with electrical signal inputs and outputs:
Input the output signal of the built-in PPG to the DUT, and then measure
the DUT output using the sampling oscilloscope.
Figure 3.2-1 When Measuring DUT With Electrical Signal Inputs and
Outputs (MP2110A-012, 021)
1. Connect the DUT input connectors to PPG1
Data
Out
and
Data
Out
using coaxial cables.
If the DUT has only one input connector, connect it to the PPG1
Data
Out
. Also, connect the coaxial terminator included as a standard
accessory to the
Data
Out
.
2. Connect the
Clk Out
(or
Sync Out
) on BERT to
Trigger Clk In
on
Scope using a coaxial cable.
When the
Sync Out
is connected, connect the supplied coaxial
terminator to
Sync
Out
.
3. Connect the DUT output connectors to
Ch A In
and
Ch B In
on the
sampling oscilloscope using coaxial cables.
When the amplitude of the input signal is greater than 400 mVp-p,
connect the attenuators to
Ch A In
and
Ch B In
for more accurate
measurement.
DUT
Summary of Contents for BERTWave Series
Page 26: ...VI...
Page 74: ...Chapter 1 Outline 1 48...
Page 166: ...Chapter 4 Screen Operation 4 24...
Page 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Page 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Page 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Page 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Page 432: ...Appendix A Specifications A 36...
Page 458: ...Appendix D Performance Test Record Form D 12...
Page 466: ...Index Index 6...