5.2 PPG/ED
Panel
5-9
5
How to Op
erate BERT
Setting Amplitude
1. Click the PPG Amplitude text box and set the amplitude voltage.
2. When the attenuators are inserted between the
Data Out
connector
and
Data
Out
connector of MP2110A and the DUTs, click the Ext
ATT text box and input the attenuation. The amplitude voltage after
passing through the attenuator is displayed.
CAUTION
When inserting attenuator at both the Data Out connector
and Data Out connector, use an attenuator with the same
amount of attenuation.
If the attenuation amounts are different, the displayed
amplitude and the amplitude output at the connector will
be different.
Summary of Contents for BERTWave Series
Page 26: ...VI...
Page 74: ...Chapter 1 Outline 1 48...
Page 166: ...Chapter 4 Screen Operation 4 24...
Page 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Page 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Page 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Page 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Page 432: ...Appendix A Specifications A 36...
Page 458: ...Appendix D Performance Test Record Form D 12...
Page 466: ...Index Index 6...