6.9 Measuring
Waveform
6-99
6
How to Operate Sampling Scope
To set Mask Margin and execute mask test
1. Click
Setup.
2. Click the Sampling Mode button to set the display Eye or Coherent
Eye.
3. Click
Sampling to set the display Run.
4. When the trace is displayed, click Auto Scale.
Check that the eye pattern is displayed in the center of the screen.
5. Click
Measure.
6. Click
Mask Test tab.
7. Click the Channel button to set the channel.
8. Click the Eye Mask Select button to select the mask.
9. Click the Align Method button to set to Zero/One/Crossing.
10. Click the Mask Test button and set the display to On.
11. Click the Test Method button to set to One Shot.
12. Click Sampling to set the display to Hold.
13. Click Update.
14 Click the mask margin text box.
15. Input the mask margin in the range of –100 to 100%.
16. The shape of the mask is changed, and the measurement result is
displayed on the screen.
Note:
When executing the Mask Test, run Auto Scale to set UI On Screen
to 2.
The Mask Test measured value is assured when UI On Screen is 2.
If UI On Screen is 3 or more, the optimum Mask position may not
be detected because the vertical scale setting is not optimum.
Summary of Contents for BERTWave Series
Page 26: ...VI...
Page 74: ...Chapter 1 Outline 1 48...
Page 166: ...Chapter 4 Screen Operation 4 24...
Page 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Page 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Page 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Page 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Page 432: ...Appendix A Specifications A 36...
Page 458: ...Appendix D Performance Test Record Form D 12...
Page 466: ...Index Index 6...