A.3 Sampling Oscilloscope
A-23
A
ppe
ndi
x A
A.3.9 Electrical Channel
Table A.3.9-1 Electrical Channel Function
Item
Specifications
Amplitude Setting
Scale
Scale Setting on the vertical axis. Both ChA and ChB can be set as follows
1 to 200 mV/div, 0.1 mV step
Offset
Offset setting on the vertical axis. Both ChA and ChB can be set as follows
–500 to 500 mV, 1 mV step
Attenuation
Function of correcting Amplitude Scale / Offset and measurement result in
conjunction with the input attenuation of the attenuator connected to Data
Input
0.00 to 30.00 dB, 0.01 dB step
Tracking*
Function to make the other channel track the active channel’s Scale, Offset
and Attenuation values.
ON/OFF
Channel Math*
Function to draw calculation results of ChA and ChB
Channel Math
Enable
ON/OFF
Define Function
Selects the calculation method
ChA + ChB
ChA – ChB
ChB – ChA
Scale
Sets scale in Vertical Axis at displaying Channel Math calculation
1 to 200 mV/div, 0.1 mV step
Offset
Sets offset in Vertical Axis at displaying Channel Math calculation
–1000 to 1000 mV, 0.1 mV step
*: Available when MP2110A-x21 is installed.
Summary of Contents for BERTWave Series
Page 26: ...VI...
Page 74: ...Chapter 1 Outline 1 48...
Page 166: ...Chapter 4 Screen Operation 4 24...
Page 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Page 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Page 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Page 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Page 432: ...Appendix A Specifications A 36...
Page 458: ...Appendix D Performance Test Record Form D 12...
Page 466: ...Index Index 6...