A.3 Sampling Oscilloscope
A-17
A
ppe
ndi
x A
*7: RJ (rms), PJ (p-p), DDJ (p-p), DCD, ISI (p-p). DDPWS and PJ
Frequency can be measured when Sampling Mode is Advanced Jitter.
*8: BER specified in TJ Measurement BER in the setting items
*9: Impossible to measure with differential signals
Table A.3.4-2 Amplitude/Time Measurement (Cont’d)
Item
Specifications
Item Selection
Selects up to 4 or 32 items from Measurement Select and displays the result
*
10
Channel, Current Value, Average, Std.Dev, Min, Max
Measuring Area
Marker
Displays the measurement area as a marker
NRZ Measure Setup
Setup items for measuring NRZ
Eye Boundary
Offset from Crossing: 0 to 1 UI (0.01 UI step)
Width:
0 to 1 UI (0.01UI step)
Rise/Fall Time
Percentage:
10-90%, 20-80%
Correction Factor: 0 to 9999.9 ps
(Correct actual Tr / Tf when inserting LPF)
Noise Measure Area Sets the position to measure noises of SNR.
Zero Level + One Level
Zero Level
One Level
PAM4 Measure Setup Setup items for measuring PAM4*
11
Sample Timing
“Track to Middle EyeTiming”:
The phase of Middle EYE is the reference
“Independent Timing”:
The phases of Upper, Middle and Lower are
independent reference in each other.
Eye Center Type
“Maximum Eye Width”: The center of the eye opening at the amplitude
where the Eye Width is the maximum is the
reference.
“Maximum Eye Height”: The phase where Eye Height becomes maximum
is the reference.
Peak Power
Hit Ratio
Sets the threshold for determining the Peak Power by the ratio to the
number of samples in 1 UI.
Pmax (Maximum Power): Position of the ratio specified to the number of
samples from the maximum value of the waveform
Pmin (Minimum Power): Position of the ratio specified to the number of
samples from the minimum value of the waveform
Peak Power: Difference between Pmax and Pmin
If 0 is specified, the difference between maximum and minimum values of
the waveform is considered Peak Power.
*10: Items 1 to 4 are available with Version 5. Items 1 to 32 are available
with Version 6 or later.
*11: Available when MP2110A-095 is installed.
Summary of Contents for BERTWave Series
Page 26: ...VI...
Page 74: ...Chapter 1 Outline 1 48...
Page 166: ...Chapter 4 Screen Operation 4 24...
Page 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Page 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Page 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Page 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Page 432: ...Appendix A Specifications A 36...
Page 458: ...Appendix D Performance Test Record Form D 12...