Chapter 6 How to Operate Sampling Scope
6-2
Measurement Procedure
6.1
The following figure shows the basic measurement procedure.
Figure 6.1-1 Basic Measurement Procedures
Start
Input clock signal to
Trigger Clk In
connector.
Input the signal to be
measured to the connector
(Ch A, Ch B, A, B, C or D) of
the Scope panel.
Set Symbol Rate.
Calibration
required?
Perform calibration.
Set to Sampling mode.
Set Clock Frequency or
Clock Divide ratio.
Eye mode?
Set pattern length.
Sets data acquired method.
Turn on the display of the
channel to which the signal
is input.
Start sampling.
Measure a waveform.
End
Stop sampling.
Adjust scale.
Track the clock
rate to BERT?
No
No
Yes
No
Yes
Yes
Save measurement result.
Use CRU clock rate?
No
Connect Trigger Clk
In connector to CRU
Out connector.
Set CRU.
Yes